Refractive Index Detector: Shodex RI-201
(Features)
· A nobel optical system (three-chamber flow cwll) provides at least twice the sensitivity of our previous detectors.
· The double temperaturefluctuations.
· The limit of detection for saccharides is approximately 2 ng.
Prenciple of New optical system measurement
in our previous optical system, the measurement light through the flow cell was refracted onlr once. the new three-chamber flow cell allows the light to be refracted wice, thereby inceasing sensitivity at least tow-times at the same optical path length. this doubles the defection degree and results in not only reduces the noise half, but reduces the drift caused by optical systems half.
(Specifications)
Product code
|
F4010105
|
Model
|
RI-201
|
Analysis
|
Flow cell type
|
3 chamber-type
|
Measuring method
|
Deflection type
|
Refractive Index range
|
1.00 to 1.75
|
Measuring range
|
0.125 to 256 micro-RIU
|
Drift
|
0.1micro-RIU/h (Pure water 1mL/min, PURGE OFF)
|
Linearity range
|
≥300micro-RIU
|
Noise
|
≤1nRIU (Pure water, response : 1.5sec)
|
Response
|
0.1 , 0.25 , 0.5 , 1 , 1.5 , 2 , 3 , 6 sec
|
Auto zero
|
Full auto zero
|
Auto zero range
|
All range
|
Off-set range
|
5micro-RIU
|
Off-set response
|
25nRIU
|
Integrator output
(Sensitivity)
|
DC 0 to 1V
(4mV/micro-RIU, 16mV/micro-RIU)
|
Cell volume
|
8micro-L
|
Flow rate
|
(Usual)
|
0.2 to 3.0mL/min
|
(Max.)
|
10mL/min (solvent : pure water)
|
Max. back pressure
|
50kPa
|
Internal volume
|
IN to Cell: ca. 80micro-L
Cell to OUT: ca. 600micro-L
All(Cell to OUT: ca.690micro-L
|
Recorder output
|
0 to 10mV/FS
|
External input
|
-
|
External output
|
(1) READY (temperature control)
(2) LEAK
(3) ERROR (ROM, RAM, PARAMETER, HOME POSITION, OVER-HEAT,
OPTI.-BALANCE, INTENSITY)
|
Temperature control
|
OFF, 30 to 50deg-C (1deg-C step)
77deg-C Temp. fuse (Double Temperature control)
|
Communication port
|
USB
|
Operational support
functions
|
None
|
Wetted materials
|
Stainless steel 316, Teflon, Quartz glass
|
Power source,
Power consumption
|
AC 100 to 240V±10%, 50/60Hz, 150VA max.
|
Demensions, Weight
|
W260 x D400 x H150 (mm), ca.12kg
|
Accessories
|
Power cable, single code, connector tube, fuse, operation manual
|
Refractive Index Detector: Shodex RI-201H
(Features)
· uses the same optical system as that of the RI-101.
· Reasonable price
(Specifications)
Product code
|
F4010106
|
Model
|
RI-201H
|
Analysis
|
Flow cell type
|
2 chamber-type
|
Measuring method
|
Deflection type
|
Refractive Index range
|
1.00 to 1.75
|
Measuring range
|
0.25 to 512 micro-RIU
|
Drift
|
0.2micro-RIU/h (Pure water 1mL/min, PURGE OFF)
|
Linearity range
|
≥600micro-RIU
|
Noise
|
≤2.5nRIU (Pure water, response : 1.5sec)
|
Response
|
0.1 , 0.25 , 0.5 , 1 , 1.5 , 2 , 3 , 6 sec
|
Auto zero
|
Full auto zero
|
Auto zero range
|
All range
|
Off-set range
|
10micro-RIU
|
Off-set response
|
50nRIU
|
Integrator output
(Sensitivity)
|
DC 0 to 1V
(2mV/micro-RIU, 8mV/micro-RIU)
|
Cell volume
|
8micro-L
|
Flow rate
|
(Usual)
|
0.2 to 3.0mL/min
|
(Max.)
|
10mL/min (solvent: pure water)
|
Max. back pressure
|
50kPa
|
Internal volume
|
IN to Cell: ca. 60micro-L
Cell to OUT: ca. 600micro-L
All(Cell to OUT: ca.670micro-L
|
Recorder output
|
0 to 10mV/FS
|
EXternal input
|
-
|
EXternal output
|
(1) READY (temperature control)
(2) LEAK
(3) ERROR (ROM, RAM, PARAMETER, HOME POSITION, OVER-HEAT,
OPTI.-BALANCE, INTENSITY)
|
Temp. control
|
OFF, 30 to 55deg-C (1deg-C step), 77deg-C Temp. fuse
|
Communication port
|
USB
|
Operational support
functions
|
None
|
Wetted materials
|
Stainless steel 316, Teflon, Quartz glass
|
Power source,
Power consumption
|
AC 100 to 240V±10%, 50/60Hz, 150VA max.
|
Demensions, Weight
|
W260 x D400 x H150 (mm), ca.12kg
|
Accessories
|
Power cable, single code, connector tube, fuse, operation manual
|
Refractive Index Detector: Shodex RI-101, 102, 104
The Shodex RI-100 series is a versatile and high sensitive RI detectors that can be used with various manufactures 'HPLC system'. It consists of a color-LCD, an automatic start-up function, and validation wizard.
(Features)
· Equipped with a color liquid crystal display, the detector monitors s in real time.
· The automatic start-up function controls the complicated operations such asblank substitution and baseline atabilization automatically.
· The validation wizard enables easy component validation.
· Improved temparature control system shortens the required warm-up time and provides stable background.
· The leak sensor automatically stops the pump in case of solvent leakage.
· External input and output terminals and RS232C communication ports can be used to connect various HPLC systems for an advanced automation.
(Specifications)
Product code
|
F4010101
|
F4010104
|
F4010102
|
Model
|
RI-101
|
RI-104
|
RI-102
|
Analysis
|
Semi-micro
|
Preparative
|
Flow cell type
|
2 chanber-type
|
Measuring method
|
Deflection type
|
Reflactive Index range
|
1.00 to 1.75
|
Drift*
|
0.2micro-RIU/h
|
0.2micro-RIU/h
|
2micro-RIU/h
|
Measuring range
|
0.25 to 512 micro-RIU
|
0.25 to 512 micro-RIU
|
2.5 to 5120 micro-RIU
|
Linearity range
|
≥600micro-RIU
|
≥600micro-RIU
|
≥6000micro-RIU
|
Noise**
|
≤2.5nRIU (response: 1.5sec)
|
≤5nRIU (response: 1.5sec)
|
≤25nRIU (response: 1.5sec)
|
Response
|
0.1 , 0.25 , 0.5 , 1 , 1.5 , 2 , 3 , 6 sec
|
Auto zero
|
Full auto zero
|
Auto zero range
|
All range
|
Off-set range
|
0 to 500mV (Same as integrator output)
|
Off-set response
|
10mV (Same as integrator output)
|
Integrator output
(Sensitivity)
|
DC 0 to 1V
(2mV/micro-RIU,
8mV/micro-RIU)
|
DC 0 to 1V
(2mV/micro-RIU,
8mV/micro-RIU)
|
DC 0 to 1V
(0.2mV/micro-RIU,
0.8mV/micro-RIU)
|
Cell volume
|
8micro-L
|
2.5micro-L
|
8micro-L
|
Flow rate
|
(Usual)
|
0.2 to 3.0mL/min
|
0.2 to 1.0mL/min
|
1.0 to 50mL/min
|
(Max.)
|
10mL/min (solvent: pure water)
|
1.0mL/min (solvent: pure water)
|
100mL/min (solvent: pure water)
|
Max. back pressure
|
50kPa
|
Internal volume
|
IN to Cell: ca. 60micro-L
Cell to OUT: ca. 600micro-L
All(Cell to OUT: ca.670micro-L
|
IN to Cell: ca. 10micro-L
Cell to OUT: ca. 355micro-L
All(Cell to OUT: ca.370micro-L
|
IN to Cell: ca. 120micro-L
Cell to OUT: ca. 510micro-L
All(Cell to OUT: ca.640micro-L
|
Recorder output
|
0 to 10mV/FS
|
EXternal input
|
Purge On/Off, Auto Zero, Marker
|
EXternal output
|
(1) READY (Automatic start-up)
(2) LEAK
(3) ERROR (OVER HEAT/LOW LIGHT INTENSITY/
NULL GLASS HOME POSITION/LOST PARAMETERS/OPTICAL BALANCE)
(Cntact capacity : DC24V 0.1A max.)
|
Temp. control
|
OFF, 30 to 55deg-C (1deg-C step), 77deg-C Temp. fuse
|
Communication port
|
RS232C
|
Operational support
functions
|
(1) Autopamtic start-up (Start Up Sequence), (2) Span/Validation Guide,
(3) Real Time Baseline Monitor
|
Wetted materials
|
Stainless steel 316, Teflon, Quartz glass
|
Powersource,
Power consumption
|
AC 100 to 240V±10%, 50/60Hz, 150VA max.
|
Demensions, Weight
|
W260 x D400 x H200 (mm), ca.13kg
|
Accessories
|
Power cable, single code, connector tube, fuse, operation manual
|
Pure water 1mL/min, PURGE OFF
Pure water, response : 1.5sec