Power Requirements
|
24 VDC @ 0.75 Amp max. (< 5 Watts consumption average)
|
|
Temperature
|
50 o C or lower when run in ambient conditions (20- 25 o C)
|
|
Vacuum Sensor Calibration Accuracy
|
Prep 80 mmHg +/- 5 mmHg Analytical 50 mmHg +/- 5 mmHg
|
|
Closed-Loop Control Setpoint
|
50 mm Hg absolute (analytical) or 80mm Hg (prep) pump runs at high RPM until near setpoint, then speed is varied to maintain a value of setpoint— load independent.
|
|
Errors Detected
|
1 – Pumpdown:
|
Prep — Unable to reach 82 mm Hg in 10 minutes. Analytical — Unable to reach 52mm Hg in 10 minutes.
|
|
2 – High Vacuum:
|
Prep — Vacuum > 110 mmHg for more than 6 min. in the running state (after pumpdown). Analytical — Vacuum > 82 mmHg for more than 6 min. in the running state (after pumpdown).
|
|
3 – Sensor Signal:
|
Sensor Failure Detected Sensor signal >800 mm Hg or sensor signal <10 mm Hg.
|
|
ANALYTICAL VACUUM PUMP TECHNICAL DATA
|
|
Air Flow (no vacuum):
|
650 SCCM @ 400 RPM; 100 SCCM @ 60 RPM
|
|
Typical Vacuum Performance:
|
50 mmHg @ (60±20) RPM (3 SCCM air flow, closed loop)
|
|
Pump-down Time:
|
<1 minute (with 4 MINI degassing channels, 50 cc total internal volume)
|
|
Pump Head Continuous Purge Air Flow Rate
|
~12 SCCM
|
|
Vacuum Flow Path Materials:
|
303 Stainless Steel, Polypropylene, PTFE, EPDM Rubber
|
|
Expected Lifetime:
|
:>5 years (continuous run @ 60 RPM)
|
|
PREP VACUUM PUMP TECHNICAL DATA
|
|
Air Flow (no vacuum):
|
650 SCCM @ 400 RPM; 100 SCCM @ 60 RPM
|
|
Typical Vacuum Performance:
|
80 mmHg @ (80±20) RPM (3 SCCM air flow, closed loop)
|
|
Pump-down Time:
|
<1 minute (with 4 MINI degassing channels, 50 cc total internal volume)
|
|
Pump Head Continuous Purge Air Flow Rate:
|
: ~30 SCCM
|
|
Vacuum Flow Path Materials:
|
303 Stainless Steel, Polypropylene, PTFE, EPDM Rubber
|
|
Expected Lifetime
|
>5 years (continuous run @ 60 RPM)
|